We understand the necessity Of providing an efficacious oxygen plant as per the Hospitals specifications. We provide tailor made, end to end solutions viz, based on location, road and ambient conditions, height above sea level without compromising on the parameters of Volume, Pressure and Purity.
This is followed with a strong backing of centralised monitoring station bringing in accountability by leveraging IOT.
The inlet air Dew point sensor protects the generator from contamination Of water. If the refrigeration dryer stops working either by operator's mistake or a malfunction, the dew point sensor sends an signal to the control system. The control systems stops the generator and avoids the contamination of the molecular sieve, until the air dewpoint comes back to the normal values.
The control system informs user about the alarm on the touchscreen through the output signal! This optional feature helps to detect the dryer problem at the beginning and saves a lot of money on molecular sieve replacement.
The PSA generator operates in cycles; the generator's pressure vessels pressurize up to maximum pressure (6—10bar) and depressurize to O in less than 5 minutes. This pressure fluctuation creates tension that must be considered while designing these vessels.
uses only vessels designed for dynamic load. The vessel is guaranteed to last 2 million pressure cycles, where standard vessels can be used only for static load.
The control system is IOT enabled, user-friendly and can be accessed over the Web Server, depending on the users hierarchy one can monitor or control and adjust the operation.